CR / DR Image Errors

 


Normally the exposure index ("S" value, EI, LgM) will indicate if the  receptor exposure level was appropriate.  However, the exposure index  may not be correctly calculated if a histogram analysis problem has  occurred.  Look at the chest area.  Is mottle present in the chest area.   If mottle is present then that is evidence that the exposure was low.  However if  the chest area does not exhibit mottle, then the mottle below the diaphragm may be the result to excessive "equalization" image  processing.  Equalization processing can make light areas darker and  darker areas lighter.  But if the lighter area represents an under exposed area then the computer is trying to build an image where the number of photons per pixel is very low and the resultant image will be noisy (mottle).
 

Click on thumbnail to see larger image

Image failed to size correctly - adult chest on 35x43 film
Agfa image plate inserted in cassette the wrong way round
Agfa image plate exposed from back of cassette then reversed in post processing and side marker replaced incorrectly (Lt sided tension pneumothorax)
Agfa image plate exposed from back of cassette
Kodak image plate exposed from back of cassette
Over aggressive black mask overlays scale on lateral ceph image
Gross under exposure increases image noise
Gross under exposure increases image noise
White line from reader error
DR reader error
 CR drop out