External Exposure Protection
X-ray machines do not produce internal radiation exposure like
radioactive materials. There are three basic protection methods for
external sources of radiation: Minimizing exposure time,
maximizing distance from the X-ray tube, and the utilization
of shielding.
Minimizing Exposure Time: Reduce "Beam-on-time"
Radiation exposure during fluoroscopy is directly proportional to
the length of time the unit is activated. Reductions can be realized
by:
- Not exposing patient while not viewing the TV image;
- Pre-planning images. An example would be to ensure correct
patient positioning before imaging to eliminate unnecessary
"panning;"
- Avoiding redundant views;
- Operator awareness of the 5-minute time notifications.
Fluoroscopy’s real-time imaging capabilities are invaluable for
guiding procedures or observing dynamic functions. However, there is
no advantage over conventional X-ray techniques when viewing static
images. Use of Last-Image-Hold features, when available, allows
static images to be viewed without continuously exposing patient and
operator to radiation.
Human eye integration time or recognition time of a fluoroscopy
image is approximately 0.2 seconds. Therefore, short "looks" usually
accomplish the same as a continuous exposure. Prolonged observation
will not improve the image brightness or resolution (Seifert 1996).
Maximize Distance
A small increase in the operator's distance from the patient can
significantly reduce the operator's exposure. Standing one step
further away from the patient can cut the physician's exposure rate
by a factor of 4 (AAPM 1998) (Figure 5-1). You should periodically
self-evaluate you personal technique to identify whether
opportunities to increase distance exist.
Figure 5-1: Benefit of Increasing Distance
Courtesy of Sorenson, 2000.

In percutaneous transluminal techniques, using the femoral
approach rather than the brachial approach yields distance benefits
to the operator (Figure 5-2).
Figure 5-2: Influence of Technique
Courtesy of Sorenson, 2000.

Substantial increases in operator distance may be realized through
remote fluoroscopy activation whenever automated contrast injectors
are used.
Many procedures require staff to intermittently interact with the
patient near the fluoroscopy system. The operator can reduce staff
exposure by delaying fluoroscopy until these activities are
completed and/or by alerting these personnel when imaging;
especially during high dose rate modes like cineangiography (Figure
5-3).
Figure 5-3: Benefit of Alerting Staff
Courtesy of Sorenson, 2000.

Room Lighting
Provisions should be made to eliminate extraneous light that can
interfere with the fluoroscopic examination. Room lighting should be
dim to enhance visualization of the image. Excessive light can
decrease the ability of the eye to resolve detail. Measures taken to
improve detail often involve increasing patient/staff exposure.
X-ray Tube Position
Fluoroscopy examinations have the smallest operator exposure when
the X-ray tube is underneath the examination table (Figure 5-3).
Whenever possible, the operator should avoid the X-ray tube side of
the table when imaging oblique or lateral images.
Figure 5-3: Benefit of Under-Table Position
Courtesy of Sorenson, 2000.

Note: The benefit is exaggerated-some operator dose
occurs on the I-I side.
I-I to Patient Air Gap
The operator must be aware of the X-ray tube-to-patient distance.
Positions closer can lead to extremely high patient exposures due to
Inverse-Square-Law effects (case study). Minimizing the air gap
between the I-I and the patient typically ensures that this distance
is maintained. Use of the separator or spacer cone can
prevent serious effects. The spacer cone is a spacer attached to the
tube housing designed to keep the patient at a reasonable distance
from the x-ray source. This is done specifically to avoid the high
skin-dose rates that can be encountered near the tube port. Spacer
cones protect patients from extremely high local exposures by making
it physically impossible to get too close to the X-ray source
(inverse-square law effects). For some X-ray machines, the spacer
cone is designed to be removable (Figure 5-3) in order to provide
more flexibility in positioning for some special surgical procedures
(e.g., portable C-arms). There is a risk of very high dose rates to
the skin surface when it is removed.
Figure 5-3: Removable Spacer Cone
Courtesy of Rauch, 2000

Reduce Air Gaps
Keeping the I-I as close to patient’s surface as possible
significantly reduces patient and operator exposures (Figure 5-4).
The I-I will intercept the primary beam earlier and allow less
scatter to operator and staff. In addition, The Automatic Brightness
Control (ABC) system would not need to compensate for the increased
X-ray tube to I-I distance caused by the air gap. The presence of an
air gap will always increase patient/operator radiation exposure
and decrease image quality.
Figure 5-4: Benefit of Reducing the Air Gap (I-I Close to
Patient)
Courtesy of Sorenson, 2000.

Care should be taken whenever the image view angle is changed
during the procedure (e.g, changing from an ANT to a steep LAO). The
I-I is often moved away from the patient while changing X-ray tube
position. Large air gaps can result if the table or I-I height
remains unadjusted.
I-I to Patient Distance Example:
After changing views, a 10-cm air gap between I-I and patient is
inadvertently maintained. What is the increase in radiation exposure
to a 20-cm thick patient positioned with the table 30 cm away from
the X-ray source, assuming the ABC compensates by increasing mA
only?
Note: mA only adjustments on ABC systems are
reasonably common.
Solution:
Assuming the air gap could have been eliminated by moving the I-I
closer, and that the brightness loss follows the inverse square law:


The brightness level with the air gap is only 69% of the zero air
gap brightness. The ABC system compensates for brightness loss by
producing 31% more X-rays. The exposure rate to the patient and
staff is subsequently increased by 31%.
Reducing air gaps between patient and I-I also reduces image blur.
Blurring of the image is caused by geometric magnification caused by
air gaps. Gaps between patient and I-I enhance geometric
magnification. The objects will appear larger with increasing gap
size. However, note that image edges are more fuzzy (Figure 5-4).
The degree of "fuzziness" will increase with increasing air gap.
Figure 5-4: I-I distance and Image Blur
Courtesy of Sorenson, 2000.

Courtesy of Sorenson, 2000.
Minimize Use of Magnification
Use of magnification modes significantly increases radiation
exposure to patient, operator, and staff (See Chapter 3).
Magnification modes should be employed only when the increased
resolution of fine detail is necessary.
Collimate the Primary Beam
Collimating the primary beam to view only tissue regions of interest
reduces unnecessary tissue exposure and improves the patient’s
overall benefit-to-risk ratio. Optimal collimation also reduces
image noise caused by scatter radiation originating from outside the
region of interest (See Chapter 3). A good rule of thumb is that
fluoroscopy images should not be totally "round" when collimators
are available for use, the collimator edges should always be visible
in the image.
Use Alternate Projections
Continuous exposure of the patient with the same projection
(point of X-ray beam entry) can cause very high skin dose to small
areas. Thus, if the point of X-ray beam (projection) entry can be
changed, the skin may be spared from the harmful effects of
radiation. While this is an effective protection method, care must
be exercised to utilize this method intelligently since longer beam
paths through the patient can cause higher patient and worker dose.
Steeply angled oblique images (e.g., LAO 50 with 30 cranial tilt)
are typically associated with increased radiation exposure since:
X-rays must pass through more tissue before reaching I-I. ABC
compensates for X-ray loss caused by increased attenuation by
generating more X-rays; Steep oblique angles are typically
associated with increased X-ray tube to I-I distances. The ABC
compensates for brightness loss caused by inverse square law effects
by generating more X-rays. Oblique views may bring the X-ray tube
closer to the operator side of the table, increasing radiation
exposure from scatter.
Operator exposure from different projections.
When possible, use alternate views (e.g., ANT, LAO with no tilt)
when similar information can be obtained (Figure 5-5). The physician
can reduce personal exposure by re-locating himself when oblique
views are taken. For example, dose rates can be reduced by a factor
of 5 when the physician stands on the I-I side of the table (versus
X-ray tube side) during a lateral projection (AAPM 1998).
Figure 5-5: Physician Exposure for a variety of Projections
Courtesy of Sorenson, 2000.

Projections with the X-ray tube neutral or tilted-away from the
operator are highlighted blue, while those tilted towards the
operator are in red. Note the decrease seen between the LAO 40
views. The caudal tilt causes the tube to be more tilted away from
the operator.
Optimizing X-ray Tube Voltage
Selection of an adequate kVp value will allow sufficient X-ray
penetration while reducing the patient’s dose rate. In general, the
highest kVp should be used which is consistent with the degree of
contrast required (high kVp decreases image contrast).
Henry Ford Hospital has many resources available (e.g., Staff
Radiologists, Medical Physicists) to assist the operator in
optimizing the fluoroscopy image while minimizing patient exposure.
Use of Radiation Shields
Use of radiation shielding is highly effective in intercepting and
reducing exposure from scattered radiation (Figure 5-6). The
operator can realize radiation exposure reductions of more than 90
percent through the correct use of any of the following shielding
options. Shields are most effective when placed as near to the
radiation scatter source as possible (i.e., close to patient).
Many fluoroscopy systems contain side-table drapes or similar types
of lead shielding. Use of these items can significantly reduce
operator exposures. Many operators have had little difficulty
incorporating their use, even during procedures requiring multiple
re-positioning of the system.
Figure 5-6: Benefit of Hanging Shield
Courtesy of Sorenson, 2000.

Ceiling-mounted lead acrylic face shields should be used whenever
these units are available, especially during cardiac procedures.
Correct positioning is obtained when the operator can view the
patient, especially the beam entrance location, through the shield.
Portable radiation shields can also be employed to reduce exposure.
Situations where these can be used include shielding nearby
personnel who remain stationary during the procedure.
Use of Personal Protective Equipment
Use of leaded garments substantially reduces radiation exposure by
protecting specific body regions. Many fluoroscopy users would
exceed regulatory limits should lead aprons not be worn. Operator
and nearby staff (within 2 meters) are required to wear lead aprons
whenever fluoroscopes are operated at Henry Ford Hospital. Due to
the poor material qualities of Leaded garments, proper storage is
essential to protect against damage (Figure 5-6). Whenever leaded
apron are required, they must be supplied and paid for by your
employer (Henry Ford Health System)
Figure 5-6: Properly Stored Leaded Garments
Courtesy of Sorenson, 2000.

Courtesy of Sorenson, 2000.
Lead aprons do not stop all the x-rays. Typically at least a 80%
reduction in radiation exposure is obtained by wearing a lead apron
(Figure 5-7). It should be noted that the apron's effectiveness is
reduced when more penetrating radiation is employed (e.g., the ABC
boost's kVp for thick patients). Two piece lead apron systems are
recommended for most users since they provide "wrap-around
protection" and distribute weight more evenly on the user. Some
aprons contain an internal frame that distributes some of the weight
from the shoulders onto the hips much like a backpack frame. So
called "light" aprons should be scrutinized to ensure that adequate
levels of shielding are provided. State of Michigan law requires
the use of 0.5 mm lead equivalent aprons.
Figure 5-7: Lead Apron Protection Efficiency
Courtesy of Sorenson, 2000.
Note that higher tube voltages sharply reduces the
shielding benefits of lead aprons. Higher tube voltages will occur
when imaging large patients or thick body portions. Also note that
light aprons (0.25 to 0.35 mm Pb) provide less protection compared
to the recommended 0.5 mm thickness.
Thyroid shields provide similar levels of protection to the
individual’s neck region. Thyroid shield use is required for
operators who use fluoroscopy extensively during their practice.
Optically clear lead glasses are available that can reduce the
operator's eye exposure by 85-90% (Siefert 1996). However, due to
the relatively high threshold for cataract development, leaded
glasses are only recommended for personnel with very high
fluoroscopy work loads (e.g., busy Radiology and Cardiology
Interventionists). Glasses selected should be "wrap-around" in
design to protect the eye lens from side angle exposures. Leaded
glasses also provide the additional benefit of providing splash
protection. Progressive style lenses for bifocal prescriptions are
available from a limited number of manufacturers.
The latex leaded gloves provide extremely limited protection.
Standard (0.5 mm lead equivalent) leaded gloves provide useful
protection to the user’s hands. However, trade-offs associated
with use of 0.5 mm leaded gloves include loss in tactile feel,
increased encumbrance and sterility. For these reasons, use of
leaded gloves is left to the operator’s discretion. To minimize
radiation exposure to the hands, the operator should:
- Avoid placing his hands in the primary beam at all times;
- Place hands only on top of the patient. Hands should never be
placed underneath the patient or table top during imaging;
- Consider using leaded gloves if hand placement within the
X-ray beam is necessary or positioned nearby for extended periods
of time.
Radiation Monitoring-Dosimeter Badges
Unlike many workplace hazards, radiation is imperceptible to
human senses. Therefore, monitoring of personnel exposed to
radiation is performed using a radiation dosimeter or "badge."
Monitoring is useful to identify both equipment problems and
opportunities for improving individual technique (ensuring radiation
doses are ALARA). Monitoring also documents the level of
occupational exposure.
The requirements for dosimetry has been determined by the Radiation
Safety Committee for each work area. These specify the types of
dosimeters issued as well as the collection frequency.
Some workers are issued a single whole body badge (black figure
icon). This whole body dosimeter should be worn on the collar
outside of any protective equipment worn (lead aprons). Readings
from this position provide an estimate of the radiation exposure to
the eyes. Dose estimates to the individual’s whole body are made
using the appropriate algorithm. Other workers are issued multiple
dosimeters. These are designed to be worn as shown (Figure 5-8):
Figure 5-8: Protective Devices
Lieto and Jackson, 2000.

Ring badge and Sterility
Infection Control has evaluated the use of ring badges in
surgical arenas. For open surgical theaters, ring badges are
contraindicated. Catheter procedures may be performed with ring
badges.
Dosimetry Practices
In order to provide an accurate estimate of personal risk, radiation
badges are to be used at all times when working with radiation. It
is also important to turn in the radiation badges on time. The
accuracy of the readings depends on the timely processing of the
dosimeter with the corresponding control dosimeters.
Absent dosimeters are taken very seriously by the institution.
Reports of which individuals have failed to properly return
dosimeters (who did not report the loss of the dosimeter to the RSO)
are sent to: the Radiation Safety Committee; the Department chairs;
the Hospital Medical Executive Committee; and the Board of the
institution. To avoid this negative attention, turn your dosimeter
in on time and promptly report the loss of a dosimeter to the
Radiation Safety Office. A new dosimeter will be issued at no cost
and your good name will be preserved.
The Radiation Safety Officer (RSO) reviews dosimetry records on a
monthly basis. Investigations of any exposure exceeding the
established standards are performed to determine whether corrective
action can eliminate or reduce exposures for all concerned. The
circumstances surrounding most cases of excessive radiation
exposures are often readily mitigated.
Radiation reports are provided annually to all monitored personnel
employed or practicing at Henry Ford Hospital. In addition, monthly
reporting of radiation exposure is available for highly exposed
fluoroscopy users. Individuals can access their personal records at
any time, and written dose estimates are provided upon request.
ALARA Philosophy
Regulatory dose limits should be viewed as the maximum tolerable
levels. Since stochastic radiation effects, such as carcinogenesis,
can not be ruled-out at low levels of exposure, it is prudent to
minimize radiation exposure whenever possible. This concept leads to
the As-Low-As-Reasonably-Achievable (ALARA) philosophy.
Simply stated, the ALARA philosophy requires that all reasonable
measures to reduce radiation exposure be taken. Typically, the
operator defines what is reasonable. The principles discussed in
this manual are intended to assist the operator in evaluating what
constitutes ALARA for his/her fluoroscopy usage.
The Henry Ford Hospital administration is committed to ensuring
that radiation exposure to its medical staff and employees is kept
ALARA. Full attainment of this goal is not possible without the
co-operation of all medical users of radiation devices.
Summary of Fluoroscopy Safety
- Keep beam ON-time to an absolute minimum!
- Always use tight collimation!
- Do not overuse the magnification mode.
- Keep the image intensifier as close to the patient as
possible, and the tube as far away from the patient as possible.
- Keep the kVp as high as possible considering the patient dose
versus image quality.
- Keep tube current (mA) as low as possible.
- Minimize room lighting to optimize image viewing.
- Do not overuse the high dose rate.
- Personnel must wear protective aprons, use shielding, monitor
doses and know how to position themselves and the machines for
minimum dose.
- Change projections angle for long procedures to minimize local
skin doses.
- Remember that the X-ray output, patient dose, and area scatter
levels increase for larger patients.