AEC 2

 


This page is copied from http://www.endlessedu.com/asrt_automatic_exposure_control.htm

    AUTOMATIC EXPOSURE CONTROL

I. INTRODUCTION

controlpanel.gif (350616 bytes)

Automatic exposure control devices can assist the radiographer in producing consistent radiographic images from patient to patient, regardless of size or presence of pathology. The advantages of this consistency are numerous and include: decreased repeat rate; decreased patient exposure; and increased department efficiency. The most important benefit being decreased repeat rate. According to Chesney's, Equipment for Student Radiographers, "Surveys conducted on a wide scale have drawn conclusions that inaccurate exposures have been the most common cause for radiographs needing to be repeated" (1994).

Although automatic timers have the potential to decrease the amount of films and increase department efficiency, this can only be accomplished if the equipment is operated by a skillful technologist. Even though it is called an "automatic" exposure device, a technologist must be very knowledable about automatic timing devices to produce high quality radiographs.

This course is designed to review the basic operation of Automatic Exposure Control devices and offer suggestions to the participant on how to utilize AEC devices to obtain optimum radiographs.

1.1 METHODS OF TERMINATING AN EXPOSURE

There are two ways that a radiographic exposure can be terminated: manually or automatically. When an examination is manually timed, the technologist sets the kVp, mA, and time. After the predetermined time has elapsed, the exposure is terminated. If the equipment is operating properly and the correct technique was used for the appropriate patient thickness, one can expect a properly exposed radiograph.

When an AEC device is used to terminate an exposure, the technologist sets the kVp and mA, but the time of the exposure is automatically determined by the machine. The AEC device differs from a manual timer because the AEC does not stop the exposure until the film has reached an appropriate density. Unlike manual timers, which simply stop the exposure after the preset time has elapsed.

A major benefit of the AEC device is its ability to consistently obtain accurately exposed radiographs, even in the presence of pathology. While manual timers terminate the exposure at the preset time, regardless of pathology or achievement of proper film density. The following example demonstrates the difference:

Two patients may come to the radiology department for chest x-rays. They both may measure 18 centimeters, one may have normal lung fields, while the other may have a pleural effusion. Since both patients measure the same thickness, the radiographer would most likely use the same technique on both patients when manually timing.

 normvspath.jpg (48246 bytes)

Chances are the radiograph of the fluid-filled lungs will be lighter than the healthy lungs and therefore it would have to be repeated. If AEC would have been used in this situation, the exposure time would have been automatically increased to compensate for the fluid in the lungs . A diagnostic radiograph would have been produced, therefore eliminating the need for a repeat radiograph.

II. AEC PHYSICS AND INSTRUMENTATION

AEC devices are common in today's radiographic equipment. When AEC devices were first introduced, they were strictly used for fluoroscopic spot films. As advances were made in technology, automatic timers were re-designed to be used in wall and table buckys. Today, AEC devices find application in general, fluorosopic and even portable radiographic equipment. Before discussing AEC devices any further, it is important to review the basic operation of an automatic timer.

2.1 TYPES OF AUTOMATIC EXPOSURE CONTROLS

The most common type of AEC devices used in today's radiographic equipment is the ionization chamber. In older equipment, the phototimer was most commonly used for the automatic timing mechanism.

Even though the ionization chamber and the phototimer operate differently, they both have the same function: convert radiation into an electrical signal which will be used to automatically stop the exposure when the film has reached the proper density.

2.2 THE IONIZATION TIMER

The ionization timer utilizes an ionization chamber, capacitor, and exposure terminating switch to automatically terminate the exposure after the film has reached the predetermined density. A brief review of how the ionization timer operates, will be beneficial at this point.

An ionization chamber is a radiation detection device that produces a small electrical current when struck by radiation. Inside the chamber are two conducting plates which are separated by air. When radiation strikes the chamber, the air inside the chamber is ionized and the electrons migrate toward the plates, thus producing a small electrical current. This electrical current is then used to charge a capacitor. When the capacitor, an electronic storage device, reaches a predetermined charge the exposure terminating switch is activated and the radiographic exposure is terminated. It is important to remember that in this situation, the exposure was terminated by sensing the amount of radiation reaching the film, and not by a preset time.

When an x-ray machine utilizes the ionization chamber as its automatic timer, three chambers are used in the configuration that is demonstrated in Figure 2.2. The ionization chambers are usually located behind the grid and in front of the cassette.

chamout1.jpg (9031 bytes)

Figure 2.2

2.3 THE PHOTOTIMER

Another type of automatic exposure device that may be used in radiographic equipment is the phototimer. The phototimer was commonly used in older x-ray equipment and consisted of a fluorescent screen, photomultiplier tube, capacitor and an exposure terminating switch. Although the components of the phototimer are different, the theory of operation is similar to that of the ionization timer which is discussed in section 2.3.

The ionization timer and phototimer both convert radiation into an electrical signal which is used to terminate the exposure when the film has reached the proper density. However, since the phototimer is a bit less sophisticated than the ionization chamber, there are a few more steps involved in the conversion of radiation into an electrical signal.

In the phototimer assembly, a fluorescent screen is placed behind the bucky. A photomultiplier tube is then placed directly behind the fluorescent screen. The fluorescent screen converts the radiation that exits from the patient and cassette into light. The photomultiplier tube then converts the light emitted from the screen into an electrical current which is used to charge a capacitor. When the capacitor is charged to the predetermined level, the exposure is terminated.

As with the ionization timer, the length of the exposure is based on the time it takes to charge the capacitor to the predetermined level and not a time set by the technologist. If a patient is larger than "average" more radiation is absorbed by the patient, and less is converted to an electrical signal. Therefore, the exposure will be longer since it will take longer to charge the capacitor to the predetermined level. Likewise, if a patient is very thin, there is minimal absorption of the beam which results in more radiation being converted into an electrical signal. This in turn will charge the capacitor more quickly and terminate the exposure more rapidly. Since the xray tube is "on " while the capacitor is being charged, it should become obvious that the length of time that it takes to charge the capacitor to the predetermined level is directly related to film density.

As mentioned earlier, the ionization timer is used more commonly in modern radiographic equipment than the phototimer. However, the term "phototiming" has become synonymous with either type of automatic exposure control.

Now that the basic operation of AEC devices have been reviewed, it is time to discuss how to properly use them.

III. TECHNOLOGIST'S DECISIONS

When using an automatic exposure control device, there are many important decisions a technologist must make in order to ensure that a diagnostic film will be obtained. The two most important are patient positioning and proper detector selection. As discussed earlier, the Phototimer or Ionization Chamber which are known as detectors collect the radiation coming from the patient and convert it into an electrical signal. For the film to have the proper density, the detector must sample the radiation coming directly from the area of interest. If the detector samples radiation from another area, the film will not have the proper density. This then explains why proper positioning is so important when when using AEC devices: Incorrect positioning will lead to a film with incorrect density.

3.1 PATIENT POSITIONING

The following diagrams will help illustrate the importance of proper patient position.

In the diagram of an incorrectly positioned shoulder, the radiograph will not have the proper density because the shoulder joint is not directly over the detector. Due to the poor positioning, a portion of the detector is completely outside of the body and will be directly exposed by the beam. This will charge the capacitor very quickly, resulting in a radiograph that is too light to fully demonstrate the shoulder.

shouldrev1.jpg (8794 bytes)

In Figure 3.1-1, the lateral spine radiograph will be too light because the detector is sampling radiation from soft tissue along with radiation emerging from the spine. Because the soft tissue is easily penetrated, a "large" electric current will be produced in the timing circuit. The capacitor will be charged quickly resulting in a radiograph that is too light because the exposure time was too short.

lumbarcelldrawing.jpg (6422 bytes)

                                        Figure 3.1-1

Without a technologist who is very knowledgeable about anatomy and positioning, automatic timers are worthless. In fact, they may actually decrease department efficiency because of the increased amount of repeat radiographs that will result if used improperly.

3.2 DETECTOR SELECTION

    detectselect.jpg (12196 bytes)

Along with proper positioning, proper detector selection also influences the operation of the AEC device. Auto timers may have one to three detectors in their circuits, most table and wall buckys have three. Because there are three, the question often arises, "Which detector should I use?" If one keeps in mind that the detector must sample radiation coming from the area of interest, the decision of which detector to select becomes an easy one.

Here are some general guidelines to follow for determining detector selection:

901btrib.gif (868 bytes) When the vertebral column is the main area of interest, the center detector should be selected.

When using AEC for joints such as the shoulder or knee, the center detector should be selected. If the outside detectors were selected, the radiograph may not have the proper density because the outer detectors may be collimated out of the field or they may detect too much radiation coming from the soft tissue. Either one of these situations will result in a radiograph that does not have the proper density.

When the pelvis is being radiographed using AEC, the two outside detectors should be selected. When a pelvis is properly positioned the two outer detectors will be directly below the ilia.

901btrib.gif (868 bytes) Detector selection for a chest x-ray is a bit more challenging because several factors influence this decision. Radiologist's preference, pathology, and surgical intervention play the biggest role in choosing which detector to use.

When the lungs are the area of interest, the right or both outer cells may be selected. The use of both outer cells for a PA chest radiograph will result in a slightly darker radiograph, since the left cell will take longer to accumulate radiation due to absorption by the heart. Therefore, when deciding between right or both outer cells for a PA chest radiograph, one should consider if the radiologist prefers darker or lighter chest films.

As a general rule, use of the left or center detector for a PA chest radiograph will result in an over exposed radiograph except in the presence of certain chest pathology or surgical intervention.

If a large pleural effusion is present in a lung and the detector over the affected lung is selected, an overexposed radiograph will result. This is explained by the fact that the fluid in the affected lung will absorb a greater amount of radiation, which in turn will result in less radiation getting to the detector. Since less radiation reaches the detector, the exposure will continue longer, and an overexposed film will result. In this situation, the cell opposite the affected lung should be selected.

If a patient has undergone a pneumonectomy, the detector on the unaffected side should be selected when using AEC for the PA chest radiograph. If the cell under the affected side was selected in this situation, an underexposed radiograph would result. Since the side of the surgical intervention would offer little absorption of the radiation, the detector would accumulate radiation very rapidly resulting in a short exposure. The film would be undiagnostic because the remaining left lung and mediastinum would not be visualized adequately.

The center cell should be selected if the mediastinum is of interest on the PA chest radiograph, and also for the lateral chest xray.

901btrib.gif (868 bytes) When using AEC for the abdomen, the technologist's choice for detector selection is once again related to the radiologist's preference for darker or lighter radiographs. Although the center, outer two, or all three detectors maybe used for the KUB; selection of the outer two cells is most technically accurate. The KUB radiograph is most commonly ordered to evaluate the soft tissue structures of the abdomen. Selection of the two outer detectors will sample the radiation coming from the soft tissue structures only, resulting in a properly exposed radiograph for the area of interest. Selection of the center cell will result in a slightly darker radiograph because the lumbar spine will attenuate a greater portion of radiation compared to soft tissue, therefore resulting in a darker radiograph. Finally, use of all three detectors will result in a radiograph having density midway between a radiograph taken with the outer detectors and a radiograph taken using only the center detector. This is explained by the fact that the detectors are sampling a portion of the radiation coming from the soft tissue and bony structures. Therefore, an electronic averaging occurs between those structures.

Figure 3.2-1 demonstrates the effect of proper cell selection and its affect on density. Radiograph A was taken with the the two outer cells selected, while radiograph B with take with the center cell selected.

kub.cellselection.jpg (60642 bytes)

                                                Figure 3.2-1

When using AEC for an upright abdomen, the center detector should be selected. Use of the outer two or all three detectors is not a good choice because of the configuration of the detectors. Since the outer two detectors are positioned higher and more laterally than the center detector, there is a chance these detectors may sample radiation coming from the base of the lungs. Because the lungs are easily penetrated, the radiographic exposure will be terminated prematurely, resulting in a radiograph that is too light. Figure 3.2-2 demonstrates that when using the outer cells for an upright abdomen, the film will lack sufficient density.

uprabd.3cells.jpg (57484 bytes)

figure 3.2-2

* It should be noted that these suggestions for cell selection are based on operational theory. Due to differences in equipment and calibration, results may vary from machine to machine.

Cassette size is another important factor that should be considered when choosing which detector to use for an exposure. When using cassette sizes that are smaller than 10" x 12", only the center detector should be used. Cassette sizes smaller than 10" x12 " have such a small area of coverage that a portion of the two outer detectors lie outside of the collimated area. If the outer detectors where selected in this situation, a portion of the radiation would never be able to reach the entire detector therefore resulting in a longer time to charge the capacitor. This would result in an unnecessarily longer exposure time and an overexposed film.

A thorough understanding of the AEC makes detector selection less threatening. Simply remember the AEC must sample radiation coming from the area of interest.

3.3 DENSITY SELECTOR

densitysel.jpg (7289 bytes)

Along with proper positioning skills and knowledge of detector selection, the radiologic technologist must understand the function of the density selector to fully utilize AEC.

When the automatic timer is installed, the capacitor in the circuit is set to terminate the exposure when it has acquired a specific charge. Remember, that as long as the capacitor is charging, the x-ray tube is producing radiation. The precise charge on the capacitor which terminates the exposure is determined by phantom studies conducted by the service engineer and radiologist input. The service engineer then adjusts the Neutral setting of the density control to correspond to an acceptable radiographic density. When a radiographic imaging unit is properly calibrated, the density selector should be kept on the "neutral" setting for the majority of procedures. If you find that you are not using the "neutral" setting for most of the exams, the unit most likely needs to be recalibrated.

There are occasions when the radiographic machine is properly calibrated, but using the neutral setting does not result in properly exposed radiographs. Some of the reasons for this will now be discussed.

Probably the biggest culprit of improper film density while using AEC is inaccurate positioning. As mentioned earlier, if the part is not properly positioned over the detector, the film will not have the correct density. The density control cannot be blamed for inadequate film density if the patient is not properly positioned!

In figure 3.3-1, the detectors have been outlined with lead wire. These radiographs demonstrate how improper positioning affects density. Radiograph B is properly positioned and therefore has sufficient density. In radiograph A, the central ray is directed too far posteriorly and in radiograph C, the central ray is directed too far anteriorly.

lumbar.celloutline.jpg (43167 bytes)

                A                             B                             C
                                       figure 3.3-1

Poor Collimation can be another cause for obtaining improperly exposed radiographs while using the AEC The detectors cannot tell the difference between the primary radiation coming from the patient and scattered radiation. Therefore, if large amounts of scattered radiation are being produced, it will be picked up by the detector(s) and cause the exposure to terminate too soon, resulting in a radiograph that has insufficient density. This radiograph will be too light, even though it was taken with density selector set at Neutral.

The radiographs in Fig. 3.3-2 demonstrate that as you increase the amount of collimation and shielding, the amount of scattered radiation reaching the detector decreases, resulting in a darker diagnostic radiograph.

3lumbarcollim.gif (32008 bytes)

Figure3.3-2

Although, little thought about, another factor that can affect the density of a radiograph while using AEC is use of the proper film/screen combination. When the automatic timer is calibrated, it is adjusted so the Neutral setting will produce a properly exposed film for a specific film/screen combination. If a cassette with a different film/screen combination is used, the film will not have the correct density because the AEC device cannot recognize changes in film/screen speed. Figure 3.3-3 demonstrates how film/screen speeds affect density.

lumbarfilmspeed.gif (48572 bytes)

Figure 3.3-3

The AEC device was calibrated to be used with the film screen combination demonstrated on the left. The film on the right has a film/screen speed combination that is sixteen times slower (extremity) than the film on the left. Both radiographs were taken at the "N" setting.

Figure 3.3-4 is another example of how film/screen combinations affect density. The machine was calibrated to be used with the film/screen combination of the radiograph on the left. The film on the right was taken with a faster film/screen speed combination, resulting in a darker film.

chest.filmspeed.jpg (49972 bytes)

         Figure 3.3-4

When using AEC devices, it is important to remember to use only the film/screen combinations that were calibrated to be used with the machine.

It is evident that the previously discussed causes of improper film density can be attributed to "user error". All too frequently AEC devices "take the rap" for improperly exposed radiographs. More often than not, AEC errors are most likely the result of poor positioning and/or collimation. However, there are situations where the technologist performed the procedure correctly but ended up with an improperly exposed film when using an AEC device at the neutral density setting. Certain patient conditions such as obesity, and ascites may affect the operation of the AEC. Because of excessive scatter produced in these situations, the AEC device may terminate the exposure prematurely.

Surgical intervention may also have an affect on the ability of the AEC device to operate properly. As discussed earlier, if a patient has had a pneumonectomy and the detector that is selected is on the same side as the pneumonectomy the image will not have the proper density.

Another factor that may influence the performance of the AEC is the presence of a prosthesis. Should the detector be directly behind a prosthesis, the radiograph will be too dark. This is easily explained by the fact that the prosthesis will prevent radiation from reaching the detector, which in turn increases the length of the radiographic exposure.

Finally, one must also consider machine failure if films taken using an AEC device do not have the proper density.

3.3.1 CHANGING THE DENSITY SETTINGS

When a radiograph needs to be repeated because of improper density, and it is not due to user error such as incorrect positioning or improper collimation, the density selector switch must be used to change the density. Since the density of a radiograph taken using AEC is determined by how long it takes to charge a capacitor to the predetermined level, the mA and time selectors no longer control the density of the radiograph as it did in manual timing. Basically, when the density selector switch is changed, the resistance in the timer circuit is also changed. Increasing the density selector by using +1 or +2 increases the resistance in the circuit, which in turn causes the capacitor to take longer to be charged to its predetermined level. Therefore, since it will take longer to charge the capacitor, the radiograph will have greater density. Likewise, choosing -1 or -2 on the density selector, deceases the resistance in the timer circuit which allows the capacitor to be charged faster than normal. This results in a film having decreased radiographic density.

At this point, it would be beneficial to review how the characteristics of kVp, mA and time change when using AEC, compared to manual timing. This will assist in understanding how to make corrections for improperly exposed radiographs obtained while using AEC devices.

901btrib.gif (868 bytes) As with manual timing, kVp still controls contrast when using AEC devices. However, the practice does exist where technologists increase kVp to increase density when an underexposed film is obtained while using AEC. This is an incorrect solution to the problem since kVp primarily controls contrast. A repeat film taken with higher kVp may appear darker than the original, however that is due to the decreased contrast of the film.

Figure 3.3.1-1 demonstrates that as you increase kVp, the films do appear darker, but as mentioned above, it is result of the change in the scale of contrast.

3knees.contrast.jpg (52835 bytes)

       Figure 3.3.1-1

Increasing kVp to produce a darker film should be avoided because the resultant change in the scale of contrast may affect interpretation of the radiograph.

901btrib.gif (868 bytes)During the use of the AEC device, the function of the mA control changes dramatically. No longer does the mA setting govern density, as it did while using manual timing. When using an automatic timer, the mA control now influences the time of the exposure. The mA selector still controls the quantity of radiation produced, however when using AEC one must remember the length of exposure is determined by the time is takes to charge the capacitor to a specified level. Therefore when using AEC devices, milliamperage influences the time of exposure because when a higher mA is selected more radiation is present. When more radiation is present, the capacitor is charged more rapidly resulting in a shorter exposure.

901btrib.gif (868 bytes)The timer control becomes inoperable when using AEC. This makes sense because the main purpose of the AEC is to terminate the exposure automatically. Certain manufacturers design their equipment so that when AEC is selected, the timer control becomes the backup timer control. This will be discussed in greater detail later in this unit.

Based on the above information, it should become obvious that when using the AEC device, density changes should only be made by using the density selector.

The density selector allows the technologist to increase or decrease the density in predetermined increments. It is a good idea to know how each density step affects the density of the radiograph so logical predictions can be made of which setting to use should a repeat be necessary. Most radiographic equipment is calibrated to that +1 corresponds to a 25% increase in density from the "N" setting and +2 corresponds to a 50% increase. The -1 setting corresponds to 25% decrease in density compared to the "N" setting, while the -2 setting corresponds to a 50% decrease. Be aware that the density settings may be adjusted for any desired increments by the service engineer.

Figure 3.3.1-2 demonstrates how density selector changes directly affect film density. The density settings are listed below each knee image. Evaluation of these images with a densitometer did indicate the following:

A 50% decrease in density from Radiograph C (Neutral)

B 25% decrease in density

D 25% increase in density

E 50% increase in density

   A                    B                   C                     D                 E

5knees.density.jpg (75920 bytes)

-2                     -1                   N                     +1               +2

IV. MINIMUM RESPONSE TIME

Occasionally a technologist may obtain an overexposed film on the -2 density setting. Even if the technologist positioned the patient properly and correctly collimated, there are times when the radiograph may still be too dark. This situation is due to a property of automatic timers known as minimum response time.

After the capacitor in the automatic timer circuit has been charged to the predetermined level, a signal is sent to terminate the exposure. Unfortunately, the exposure is not terminated instantaneously. Anywhere from three to thirty milliseconds may elapse before all the electronics and relays actually stop the exposure. This "lag" time is referred to as Minimum Response Time (MRT). MRT poses a problem to the radiographer when the MRT is longer than the time required for the exposure. In situations where the MRT is longer than the required exposure time, the radiograph will be too dark regardless of the density setting. An example will help clarify this concept:

The minimum response time of a machine is 5 milliseconds. An elderly woman with emphysema requires a chest xray. As a result of her condition, the woman's chest x-ray will only require a 2 millisecond exposure. Even though the required time for the exposure is only 2 milliseconds, the actual exposure time will be 5 milliseconds because it takes the machine a minimum of 5 milliseconds to terminate an exposure. Therefore, the radiograph will be dark when using the "N", "-1" or "-2" settings of the density selector because the machine simply cannot shut off in two milliseconds.

Problems with MRT are more common in older equipment since the MRT of these units can be as long as 30 milliseconds. Newer equipment is not faced with MRT problems as frequently as older units because advances in technology have reduced MRT's to as little as 1 millisecond.

4.1 SOLUTIONS

Problems associated with long MRT's can be solved by the use of a lower mA setting, decreased kVp, or simply using manual timing. The use of a lower mA setting is the preferred method to correct MRT problems.

When using a lower mA setting, the required time for the exposure is increased (less radiation is present, therefore a longer exposure). The goal is to make the present exposure time longer than the MRT of the machine. By making the the exposure time longer than the MRT, the automatic timer can now terminate the exposure at the proper time which in turn will result in a film with the correct density.

If the mA cannot be selected while using AEC (i.e. portable units equipped with AEC), the kVp can be lowered which will lead to an increased exposure time. Lower kVp settings result in decreased x-ray production, therefore it will take longer to charge the capacitor to appropriate level resulting in a longer exposure. However one must be careful when using kVp to change exposure time because of the affect on contrast.

Finally,the technologist should always consider one definite way to solve problems associated with the MRT of automatic timer and that is to use manual timing.

V. THE BACK-UP TIMER

One other feature that must addressed when discussing AEC devices is the back-up timer. In order to minimize errors which lead to repeat radiographs, the technologist must be fully aware of the purpose and operation of the back-up timer.

The back-up timer is a safety device which prevents the patient from receiving an excessive dose of radiation should the automatic timer fail due to mechanical or operator errors.

The back-up timer automatically terminates a phototimed exposure if it exceeds 600mAs (Federally mandated). Some radiographic units allow the operator to set the back-up time, while others are fixed at 600 mAs. If the radiography machine allows the back-up time to be adjusted, a good rule of thumb to follow is to set the back-up timer for two to three times the estimated mAs of the exposure. It is poor practice to always set the back-up timer to the maximum level because if the automatic timer fails, the patient will be unnecessarily exposed to the radiation produced by a 600 mAs exposure before radiation production is actually terminated. If the machine allows the user to select the back-up time for the exposure, the mA selector and timer control automatically become the back-up timer selector.

5.1 REASONS WHY THE BACKUP TIMER IS ACTIVATED

The are several reasons why the back-up timer may be activated and the exposure terminated. Excessively large patients or pathologic conditions may activate the back-up timer. However, it is reasonably safe to say that the main reason the back-up timer is activated during an exposure is a careless technologist. At this point, it will be helpful to review the most common errors that cause the back-up timer to be activated:

901btrib.gif (868 bytes)Wrong Detector Selected: If a patient is being radiographed on the table and the chest board detectors are accidently selected, the film will be too dark and the back-up timer will be activated. This occurs because the selected detectors on the chest board did not receive any radiation so the exposure continued until it was terminated by the back-up timer. Unfortunately, should this occur, the patient is exposed by the primary beam much longer than is necessary.

901btrib.gif (868 bytes)Wrong Tube Selected: If a procedure room has two radiographic x-ray tubes and the tube over the bucky is not the one that is energized, the exposure will continue until it is terminated by the back-up timer. This is explained by the fact that tube over the selected detector is not producing any radiation, therefore the exposure would continue indefinitely if not terminated by the back-up timer.

901btrib.gif (868 bytes)Incorrect Tube/Bucky Alignment: If the x-ray tube is not centered to the bucky, the back-up timer may be activated. Since the detectors are located in the bucky assembly, they cannot sense any radiation if they are not aligned with the tube. Therefore, the exposure will continue longer than necessary.

901btrib.gif (868 bytes)Incorrect Back-up Timer Setting: If the back-up timer is set for too short of a time, the back-up timer will terminate the exposure prematurely, resulting in an underexposed radiograph. This happens most frequently if the previous exam used a short manual time and the back-up timer was not set before the exposure was made.

An important point to remember about the back-up timer is that it is a safety feature which is used to prevent the patient from receiving excessive radiation due to machine failure. More often than not, the back-up timer is activated because of technologist error instead of equipment failure. Unfortunately, when the back-up timer is activated the patient has already been improperly exposed to radiation, and a repeat radiograph will be necessary resulting in additional unnecessary radiation exposure to the patient. Therefore, a radiologic technologist should always make sure all the automatic timer controls are properly set before the radiographic exposure is made.

VI. SUMMARY

The development of automatic exposure control has brought with it the possibility of decreased repeat rates and increased productivity. However, this is only possible with a technologist who fully understands all of the steps that are necessary to produce a diagnostic film while using AEC. Technologists who do not have a complete understanding of AEC may actually find themselves with an increased repeat rate, since its use requires precise positioning and thorough knowledge of the equipment. Perhaps a better term for automatic exposure control should be "assisted" exposure control because in reality the technologist does assist the automatic timing device in producing a diagnostic image.

The following is a summary of the steps to success when using AEC:

901bdmdb.gif (887 bytes)    The performance of any automatic timer is dependent on the knowledge and skill of the technologist.

901bdmdb.gif (887 bytes)    Positioning is crucial when using AEC devices.

901bdmdb.gif (887 bytes)    Choose the detectors that are directly below the dominant area of interest.

901bdmdb.gif (887 bytes)    The detectors cannot differentiate between secondary and scattered radiation. Therefore, proper collimation is very important.

901bdmdb.gif (887 bytes)    The back-up timer is a safety feature and should be set approximately 2-3x greater than the estimated mAs of the exposure.

901bdmdb.gif (887 bytes)    Density changes should only be made by changing the density selector.

901bdmdb.gif (887 bytes)    When using AEC, kVp changes should only be made to vary contrast.

901bdmdb.gif (887 bytes)    To decrease the density of dark films due to long minimum response times, lower the mA.

901bdmdb.gif (887 bytes)    If used properly, automatic timers can decrease repeat rates and increase department efficiency.


 

VII. REFERENCES/SUGGESTED READINGS

Carter, P.H., ed. Chesneys' Equipment for Student Radiographers. Cambridge, MA: Blackwell Scientific Publication, Inc., 1994.

Cornuelle, A.G. and Gronefeld, D.H. Radiographic Anatomy Positioning: An Intergrated Approach. Stamford, CT: Appelton & Lange, 1998.

Curry, T.S.; Dowdey, J.E.; Murry, R.C. Christensen's Physics of Diagnostic Radiology 4th ed. Malvern, PA: Lea & Febiger, 1990.

Thomspson, M.A.; Hattaway, M.P.; et al. Principles of Imaging Science and Protection. Philadelphia, PA: W.B. Saunders Company, 1994.

© 1998 Endless Education Ventures.

 

Useful Link :http://www.endlessedu.com/asrt_automatic_exposure_control.htm