Collimator

 


Typical basic collimator specification


EXTERNAL COVER IN A.B.S. PLASTIC.
Multilayer, square field X-ray collimator for stationary units.

The collimator enclosure is constructed of double steel, lead-lined walls for maximum X-ray protection .

X-ray field size is limited by 6 pairs of shutters four of which are lead-lined. Two pairs of shutters are positioned near the focus, two near the entrance window and two near the exit window of the X-ray beam from the collimator. The shutters are controlled by two knobs located on the front panel. Besides the minimum in-built filtration, three variable filters may be added manually.

SPECIFICATIONS:

External adjustment of mirror angulation.
Additional filtration:
0,1mm copper in addition to the 1mm Al. Support
0,2 mm copper in addition to the 1mm Al support.
1mm aluminium in addition to the 1 mm Al support
High luminosity provided by a quartz iodide lamp.
Timer limiting projection lamp exposure time to 30 seconds, adjustable, thus extending lamp life and preventing overheating.
150 kVp radiation shielding.
Minimum inherent filtration 2mm aluminium equivalent. (0,3mm and 1mm on request)
Continuous film coverage from 0 x 0 to 43 x 43 cm,
±1% FFD, at 90cm FFD (SID).

Notes
 

When it comes to radiation protection, the collimater plays an important role:

It is used to narrow the radiation field to a size needed for the examination at hand. For this it is equipped with sets of lead plates providing either a round or a square-shaped radiation field.

These collimating plates are either motorized or operated manually. In automatic mode, the image-receiver size is detected, and the collimating plates are operated accordingly

 

Diagrams of collimator with (1) no light beam    (2) Light beam
    