Dose Area Product Meter (DAP)

 


Dose-Area-Product (DAP) meters are large-area, transmission ionization chambers and associated electronics. In use, the ionization chamber is placed perpendicular to the beam central axis and in a location to completely intercept the entire area of the x-ray beam. The DAP, in combination with information on x-ray field size can be used to determine the average dose produced by the x-ray beam at any distance downstream in the x-ray beam from the location of the ionization chamber. The use of DAP is discussed further later.(2)
A recent modification of the ionization chamber design used in a DAP meter has resulted in an instrument that measures both DAP and the dose delivered by the x-ray beam. This design effectively combines data from a small ionization chamber that is completely irradiated by the beam and independent of the collimator adjustments with the conventional DAP meter.

Some fluoroscopic and radiographic systems have dose-area product (DAP) meters. DAP meters measure the radiation dose to air, times the area of the x-ray field. The relationship between DAP and exposure-area product (EAP) is essentially a single conversion factor that relates dose to exposure. EAP is expressed in roentgen-cm2 (R-cm2) and DAP is expressed ingray-cm2 (Gy-cm2).

How is DAP measured?
An ionization chamber larger than the area of the x-ray beam is placed just beyond the xray collimators. The DAP ionization chamber must intercept the entire x-ray field for an accurate reading, one proportional to the EAP. The reading from a DAP meter can be changed by altering the x-ray technique factors (kVp, mA, or time), varying the area of the field, or both. If the chamber area is larger than that of the collimators, as the collimators are opened or closed the charge collected will also increase or decrease in proportion to the area of the field. For example, a 5 x 5 cm x-ray field with an entrance dose of 1 mGy will yield a 25 mGycm
2 DAP value. If the field is increased to 10 x 10 cm, with the same entrance dose of 1 mGy the DAP increases to 100 mGy-cm2, which is 4 times the DAP for the 5 x 5 cm field.

      

DAP meter Sensor                        DAP meter display with print out  (Vertec)

Why DAP?
Dose-area product is relatively easy to measure. DAP meters have been around for many years, and were actually used in the 1964 and 1970 U.S. X-ray Exposure Studies. Advocates of DAP meters contend that the DAP is a better indicator of risk than entrance dose alone, since DAP incorporates the entrance dose and field size. DAP has been shown to correlate well with the total energy imparted to the patient, which is related to the effective dose and therefore to overall cancer risk.

Are there problems with DAP?
There are several problems with the use of the DAP value. The configuration of the DAP meter may introduce a bias to the DAP value. For example, if any material is placed between the meter and patient, the patient will receive less than what is implied by the displayed DAP value. For an undertable fluoroscopy system this can be the tabletop and pad. Consequently, the use of DAP to estimate skin entrance exposure or skin dose is complex and should only be attempted by a qualified medical physicist. This is particularly true for fluoroscopic procedures where multiple beam directions, source-skin distances, and field sizes may be used. DAP meters are difficult to calibrate and maintain. Large changes in the DAP meter response can occur over time, particularly if meters are adjusted for couch transmission factors. Calibration should be done in the field after any changes that might alter the DAP and at least annually.

2. Dose-Area-Product Problems in usage

Up to a decade ago, radiological patient safety concerns were focused on stochastic risk. Monitoring and managing stochastic risk requires estimates of the effective dose delivered to the patient. There is no need for real-time feedback. DAP is defined as the integral of dose across the X-ray beam. Therefore DAP includes field non-uniformity effects such as anode-heel-effect, and the use of semi-transparent beam-equalizing shutters (lung shutter). DAP is easy to measure. The simplest method is to place a transmission full-field ionization chamber in the beam between the final collimators and the patient. DAP may also be obtained by calculation. Data is accumulated during fluoroscopy, fluorography, and radiography. Assuming that the incident beam is totally confined to the patient, the recorded value essentially provides an upper limit on the X-ray energy absorbed by the patient (i.e. there is no transmission or scatter). DAP’s ability to estimate stochastic risk is degraded because of the lack of dose distribution information within the patient. The best that one can do is to assume an average weighting factor for all the tissues at risk. This may lead to an over or under estimate of risk in certain cases. As an example, it does not account for the differential risk of breast cancer from an AP or a PA projection. DAP rate and cumulative DAP can easily be displayed in real-time. The primary utility of DAP rate is in a teaching situation. Scattered dose rate at any place in the lab is more or less proportional to DAP rate. The trainee can be shown that reducing DAP rate reduces his or her personal exposure rate. The effect of different control options (e.g., collimation, zoom mode) on DAP rate can be demonstrated. Cumulative DAP does not provide a direct indication of the possibility of skin injury. The same DAP is observed with large fields and low skin doses as with small fields and high skin doses. Exceeding skin tolerance is more likely in the latter case. However, reasonable entrance field size estimates can be made for many procedures. These estimates are dependent on factors such as equipment configuration, patient size, and operator technique. Once known, the nominal field size can be used to obtain an estimate of skin dose. Rules-of-thumb can be established to make this conversion for typical procedures. DAP provides no information regarding the spatial distribution of the entrance beam around the patient’s skin. It produces an overestimate of the possibility of exceeding the deterministic threshold when there is significant beam movement during the procedure.

Summary:
DAP meters are valuable quality control tools for monitoring changes in equipment and procedures. DAP does not represent radiation dose per se, and use of a DAP meter to determine patient dose should only be attempted by a qualified medical physicist. DAP meters need to be recalibrated on a regular basis
at least annuallyto maintain adequate accuracy.

 

Useful Link : http://www.vertec.co.uk