|
Dose-Area-Product (DAP)
meters are large-area, transmission ionization chambers and
associated electronics. In use, the ionization chamber is placed
perpendicular to the beam central axis and in a location to
completely intercept the entire area of the x-ray beam. The DAP, in
combination with information on x-ray field size can be used to
determine the average dose produced by the x-ray beam at any
distance downstream in the x-ray beam from the location of the
ionization chamber. The use of DAP is discussed further later.(2)
A recent modification of the ionization chamber design used in a DAP
meter has resulted in an instrument that measures both DAP and the
dose delivered by the x-ray beam. This design effectively combines
data from a small ionization chamber that is completely irradiated
by the beam and independent of the collimator adjustments with the
conventional DAP meter.
Some fluoroscopic and radiographic systems have
dose-area product (DAP) meters. DAP meters measure the radiation
dose to air, times the area of the x-ray field. The relationship
between DAP and exposure-area product (EAP) is essentially a single
conversion factor that relates dose to exposure. EAP is expressed in
roentgen-cm 2
(R-cm2)
and DAP is expressed ingray-cm2
(Gy-cm2).
How is DAP measured?
An ionization chamber larger than the area of the x-ray beam is
placed just beyond the xray collimators. The DAP ionization chamber
must intercept the entire x-ray field for an accurate reading, one
proportional to the EAP. The reading from a DAP meter can be changed
by altering the x-ray technique factors (kVp, mA, or time), varying
the area of the field, or both. If the chamber area is larger than
that of the collimators, as the collimators are opened or closed the
charge collected will also increase or decrease in proportion to the
area of the field. For example, a 5 x 5 cm x-ray field with an
entrance dose of 1 mGy will yield a 25 mGycm 2
DAP value. If the field is
increased to 10 x 10 cm, with the same entrance dose of 1 mGy the
DAP increases to 100 mGy-cm2,
which is 4 times the DAP for the 5 x 5 cm field.

DAP meter Sensor
DAP meter display with print out (Vertec)
Why DAP?
Dose-area product is relatively easy to measure. DAP meters have
been around for many years, and were actually used in the 1964 and
1970 U.S. X-ray Exposure Studies. Advocates of DAP meters contend
that the DAP is a better indicator of risk than entrance dose alone,
since DAP incorporates the entrance dose and field size. DAP has
been shown to correlate well with the total energy imparted to the
patient, which is related to the effective dose and therefore to
overall cancer risk.
Are there problems with DAP?
There are several problems with the use of the DAP value. The
configuration of the DAP meter may introduce a bias to the DAP
value. For example, if any material is placed between the meter and
patient, the patient will receive less than what is implied by the
displayed DAP value. For an undertable fluoroscopy system this can
be the tabletop and pad. Consequently, the use of DAP to estimate
skin entrance exposure or skin dose is complex and should only be
attempted by a qualified medical physicist. This is particularly
true for fluoroscopic procedures where multiple beam directions,
source-skin distances, and field sizes may be used. DAP meters are
difficult to calibrate and maintain. Large changes in the DAP meter
response can occur over time, particularly if meters are adjusted
for couch transmission factors. Calibration should be done in the
field after any changes that might alter the DAP and at least
annually.
2. Dose-Area-Product Problems in usage
Up to a decade ago, radiological patient safety
concerns were focused on stochastic risk. Monitoring and managing
stochastic risk requires estimates of the effective dose delivered
to the patient. There is no need for real-time feedback. DAP is
defined as the integral of dose across the X-ray beam. Therefore DAP
includes field non-uniformity effects such as anode-heel-effect, and
the use of semi-transparent beam-equalizing shutters (lung shutter).
DAP is easy to measure. The simplest method is to place a
transmission full-field ionization chamber in the beam between the
final collimators and the patient. DAP may also be obtained by
calculation. Data is accumulated during fluoroscopy, fluorography,
and radiography. Assuming that the incident beam is totally confined
to the patient, the recorded value essentially provides an upper
limit on the X-ray energy absorbed by the patient (i.e. there is no
transmission or scatter). DAP’s ability to estimate stochastic risk
is degraded because of the lack of dose distribution information
within the patient. The best that one can do is to assume an average
weighting factor for all the tissues at risk. This may lead to an
over or under estimate of risk in certain cases. As an example, it
does not account for the differential risk of breast cancer from an
AP or a PA projection. DAP rate and cumulative DAP can easily be
displayed in real-time. The primary utility of DAP rate is in a
teaching situation. Scattered dose rate at any place in the lab is
more or less proportional to DAP rate. The trainee can be shown that
reducing DAP rate reduces his or her personal exposure rate. The
effect of different control options (e.g., collimation, zoom mode)
on DAP rate can be demonstrated. Cumulative DAP does not provide a
direct indication of the possibility of skin injury. The same DAP is
observed with large fields and low skin doses as with small fields
and high skin doses. Exceeding skin tolerance is more likely in the
latter case. However, reasonable entrance field size estimates can
be made for many procedures. These estimates are dependent on
factors such as equipment configuration, patient size, and operator
technique. Once known, the nominal field size can be used to obtain
an estimate of skin dose. Rules-of-thumb can be established to make
this conversion for typical procedures. DAP provides no information
regarding the spatial distribution of the entrance beam around the
patient’s skin. It produces an overestimate of the possibility of
exceeding the deterministic threshold when there is significant beam
movement during the procedure.
Summary:
DAP meters are valuable quality control tools for monitoring
changes in equipment and procedures. DAP does not represent
radiation dose per se, and use of a DAP meter to determine patient
dose should only be attempted by a qualified medical physicist. DAP
meters need to be recalibrated on a regular basis —at
least annually—to
maintain adequate accuracy.
|