Exposure Factors

 


Note this article is primarily about film and screen imaging

With CD and DR, Kv has a significantly less effect on image contrast and mAS is primarily concerned with image quality

Kv / Kvp Potential difference between film and anode
The energy (you can consider this the penetrating power) of the x-ray beam is controlled by the voltage adjustment. This control usually is labelled in keV (thousand electron volts) and sometimes the level is referred to as kVp (kilovoltage potential). Do not be confused by the different terminology, just remember there is a control by which the difference in potential between the cathode and anode can be controlled. The higher the voltage setting, the more energetic will be the beam of x-ray. A more penetrating beam will result in a lower contrast radiograph than one made with an x-ray beam having less penetrating power. It is probably obvious that the more energetic the beam, the less effect different levels of tissue density will have in attenuating that beam.
The generator waveform if is not constant potential (medium frequency etc) will affect the effective Kv.
 

mA Tube Current
The second control of the output of the x-ray tube is called the mA (milliamperage) control. This control determines how much current is allowed to flow through the filament which is the cathode side of the tube. If more current (and therefore more heating) is allowed to pass through the filament, more electrons will be available in the "space charge" for acceleration to the target and this will result in a greater flux of photons when the high voltage circuit is energized. The effect of the mA circuit is quite linear. If you want to double the number of "x" photons produced by the tube, you can do that by simply doubling the mA. Changing the number of photons produced will affect the blackness of the film but will not affect the film contrast.
 

 S Time
The third control of the x-ray tube which is used for medical imaging is the exposure timer. This is usually denoted as an "S" (exposure time in seconds) and is combined with the mA control. The combined function is usually referred to as mAs or milliampere seconds so, if you wanted to give an exposure using 10 milliampere seconds you could use a 10 mA current with a 1.0 second exposure or a 20 mA current for a 0.5 second exposure or any combination of the two which would result in the number 10. Both of these factors and their combination affect the film in a linear way. That is, if you want to double film blackness you could just double the mAs.
 

The X-Ray beam
The x-ray beam has two main properties you need to understand.

1) Beam QUALITY is the ability of the beam to penetrate an object, its all about the penetrating power of the x-ray photons, this is controlled by the KV control.

2) Beam INTENSITY this is the number of x-ray photons in the beam and is principally controlled by the mAS

But note as you increase the KV not only does the QUALITY harden (more penetrating) but you do actually get more photons so INTENSITY increases too.

Putting it all together the exposure
Any radiographic subject has a minimum Kv required for the x-ray photons penetrate the most dense part of the subject, the most radiographicaly dense part of the subject will depend upon what the part is chemicaly composed of (Atomic number) and its thickness (remember linear attenuation coefficients and HVL!?)

The thicker the subject the more absorption of x-rays so the thicker the part the more mAS you require.

In theory the more Kv you use the less the contrast of the image will have
However in practice film screen / processing conditions affect contrast much more

In practice it is not as simple as this as scatter is produced which is not image forming but adds density to the film and needs to be controlled, if you remember all those complex diagrams about interactions of x-rays with matter you will realise the amount and direction of scatter depends on the Kv and the material absorbing the x-rays.
 
Image 1  The Kv is too low the femoral condyle is under pentrated you cannot see the bone trabecualr patterns. the contrast is too high to demonstrate all the soft tissues.
Image 2 Much better the all the subject is penetrated and all the soft tissues are visible
A well exposed abdomen image demonstrating all the soft tissue structures.
A good chest image the mediastinum is pentrated the image is exposed well demonstrating the bones and soft tissues.
Under penetrated
OK
Under penetrated
Too much mAS
Too Little mAS

A few myths
Changing the Kv by  2 or 3 makes almost no perceptable image change!
Adding 10 Kv does not double the image density
Exposure factors are an exact science !
(the image you produce must satisfy the radiologist who interprets the image - not all radiologists like the same penetration / density / contrast for the same body part)


Image Contrast
Here, we need to spend a little more time discussing the issue of radiographic contrast. This is an important concept because image contrast plays a critical part in the interpreter's ability to detect abnormalities which are only slightly different from the density of the surrounding material. It is not possible to say what is the optimal contrast (or the optimal radiographic technique) for all situations. Different body parts have different inherent tissue contrast. This can be illustrated by using the extreme examples of the chest and the breast. In the chest, there is good inherent tissue contrast with densities ranging all the way from bone at the high end to air at the low end. On the other hand, the breast is inherently very low in tissue contrast only containing structures which are water density (glandular material or tumor) or fat density. For the moment, we will disregard small calcifications which are really not normal structures. Because of this difference in inherent tissue contrast, we would be likely to use a very low contrast radiographic technique for the chest because we have good tissue contrast. Conversely we would be likely to use a very high contrast technique for the breast because the breast has minimal, inherent tissue contrast.

Remember, image contrast is controlled by the energy of the "x" photon beam. Therefore, high kV techniques result in low contrast images (the assumption is always made that the image will have approximately the same average film density so if kV is increased, there must be a compensation in mAs to keep film density constant). To increase image contrast in situations where there is low tissue contrast, a low kV, high mAs technique should be used. This is obvious for mammography but you should also remember this possibility for other special situations such as looking for low-density foreign bodies embedded in soft tissue. To improve film contrast for mammograms we would need to use a very low energy x-ray beam. Mammograms are frequently done with beams in the 25 keV range. For the chest x-ray, we would like to use a low contrast technique which requires a relatively high-energy beam. Chest x-rays are frequently done with beam energies above 100 keV. You should understand that for similar film densities, the high KV technique usually results in lower patient radiation exposure. Think about this long enough to clearly understand why less radiation is absorbed in the patient when a high-energy beam is used.

Grids
One of the problems in getting a sharply defined image in clinical radiology is the presence of scattered or secondary radiation. These photons are created in the body of the patient or closely surrounding objects by the interaction of that material and the primary "x" photons coming from the x-ray tube. Several possible interactions occur in the diagnostic energy range. At relatively low energies, the photoelectric effect is probable. The photoelectric effect is actually the desirable, photon/tissue interaction because there is complete absorption of the photon with no production of a secondary photon. The more common tissue interaction at the photon energies used for the majority of clinical procedures is called the Compton effect or coherent scattering. In this interaction, a secondary photon is produced at the site of interaction. The secondary photon will always have lower energy than the primary photon and will be going in an altered direction. These secondary photons, if allowed to reach the film, will actually produce erroneous information by recording gray tone variation (and therefore indicating relative tissue densities) at some distance from the site at which the photon/tissue interaction actually occurred. The net result of allowing a significant number of secondary photons to reach the film is a reduction in image sharpness. There will always be a loss of spatial resolution.

Several methods have been devised to reduce the problem of scattered radiation. The simplest and most direct is to simply limit the field of exposure. If a small image area is adequate to make the clinical diagnosis, the image area should be "coned down" to that small size. For instance, if you want to image the gallbladder, you will get a much sharper picture if you bring the shutters down to include an area only the size of the gallbladder instead of including the entire upper abdomen on the image. Just remember that the smaller the area of the x-ray beam the fewer scattered photons you will produce.

In the typical clinical imaging situation, the most common method of reducing scatter is to use a radiographic grid. The grid looks like a flat metallic plate the size of the x-ray film if you look at it directly. However, it is more complicated than that. It actually is composed of alternating radiopaque (lead) and radiolucent (aluminum) strips. These are arranged on edge, sort of like looking at the strips of a venetian blind which is arranged to let light come between the strips. The edge of these strips is turned towards the source of x-rays and in the most commonly used grid, the focused grid, the anglulation of the strips is arranged to match the divergence of the x-ray beam.

This arrangement of the radiographic grid will give the highest probability for primary "x" photons passing between the lead grid strips and reaching the film, while the off-focus or secondary photons are likely to interact in the lead strips and never reach the film.

The use of this radiographic grid will greatly improve image sharpness when a relatively thick body part is being imaged. Unfortunately, there is always a trade off. Since the grid does stop some of the photons which would contribute to film blackening, if you just add a radiographic grid without changing the tube settings, the film will be greatly underexposed. If you decide to use a grid, you will have to increase the number of photons produced by the x-ray tube in order to get the correct film exposure. This will result in giving the patient increased radiation exposure. Remember, the position of the grid is between the patient and the film.

The third method of reducing scatter or at least reducing the probability that scattered photons will reach the film is to use an air gap. This is infrequently used in clinical radiography but can still, sometimes be used to an advantage particularly when magnification of the image might be helpful. Ordinarily we would have the film positioned as close to the patient's body as possible for the radiography of any body part. With an air gap technique, the film is moved several inches away from the patient's body. That separation, (because secondary photons are likely to be lower energy and moving at a greater angle than primary photons) will result in a decreased probability of the secondary photon hitting the film. From the diagram below, you will be able to understand that creating the air gap will also result in magnifying the radiographic image. Remember the x-ray beam is produced from almost a point source and it diverges as it goes towards the patient.
 

Useful Link http://www.med.sc.edu:1000/2prod&useab.htm