Split phantom test and how is it used to test sensitometric
differences between different emulsions or batches of film? (From
the Kodak website)
A split phantom test should be performed to radiographically
determine relative speed differences between two different boxes of
film, one of which is suspected of being much faster or slower than
the film in current use for either clinical films or for processor
quality control. Speed comparisons made using a sensitometer may not
accurately reflect the differences in speed between two films
exposed by light from an intensifying screen.
The procedure is as follows:
-
- Assemble the tools that are needed for the test:
- A phantom used for mammography quality control testing
- The 18 x 24 cm mammography cassette normally used for the
phantom test
- A piece of cardboard from the film box cut in half to use as
a guide
- A pair of scissors
- A lead pencil
The mammography x-ray unit and the processor will also be used
for this test.
- In the darkroom (in total darkness to reduce any additional
density added to the films due to long safelight exposure) cut a
sheet of film from the current or "normal" box in half by using
the cardboard as a guide. (This can be done by lining up the 18 cm
edges of the cardboard and film so that the film is closest to the
countertop and the cardboard half is on top. Be careful cutting
the film in the dark.)
- Place the film--emulsion side up--in the cover of the opened
cassette with the film on the right side and the cut edge toward
the right edge of the cassette; use a lead pencil to mark the
corner "N" for normal.
- Cut a sheet of film from the "suspect" box in half by using
the cardboard as a guide.
- Place the film--emulsion side up--in the cover of the opened
cassette with the film on the left side and the cut edge toward
the left edge of the cassette; use the lead pencil to mark the
corner "S" for suspect.
- Before closing the cassette, make sure the film edges in the
center of the cassette are directly adjacent to one another and
not overlapping.

- Place the cassette with the two film halves in the grid of the
mammography x-ray unit.
- Place the phantom on top of the grid in the standard location
used for mammography quality control testing.
- Position the photocell beneath the center of the phantom
(standard location), assuming the phantom exposure is always made
using the phototimer.
- Select the same technique factors usually employed when
imaging the phantom (same kVp, etc.).
- Make the exposure and immediately process the two film halves
in the same manner (e.g., emulsion side up and on the right side
of the processor).
- Use a densitometer to take two optical density readings in the
center of the phantom, just to the right and left of the cut edges
(one on the "normal" and one on the "suspect" film).
- Calculate the density difference by subtracting the optical
density value of the "suspect" film from the optical density value
of the "normal" film.
If the density difference is a negative value and the "suspect"
film is darker than the "normal" film, the "suspect" film is
faster. If the density difference is a positive value and the
"suspect" film is lighter than the "normal" film, the "suspect"
film is slower.
According to the American College of Radiology in
Recommended Specifications for New Mammography Equipment
:
- "A density difference of 0.30 between any two films of the
same type from the same manufacturer, exposed and processed
together, is a reasonable maximum to be expected from
manufacturing variability for films of roughly the same age and
storage conditions."
- "If the difference between the two film densities exceeds 0.30
at a density of approximately 1.25, then the film supplier should
be contacted to determine the source of the problem."
Note that a difference of 0.30 at a density of
approximately 1.25 may translate into a bigger difference for
clinical films exposed at a greater optical density. For example,
high-contrast mammography films, such as KODAK MIN-R 2000 Film, are
frequently exposed at an optical density between 1.50 and 2.00 in
order to maximize contrast. The density difference at this optical
density level will be greater due to the increased contrast.
|