Radiation protection in chest radiography

 

The results below are from an experiment to determine the best position of lead protection during chest imaging to minimise gonad dose.

Data from a lab experiment looking at the posterior vs. anterior shielding 'debate'. Bottom line, the vast majority of the scatter gonads (and by extension, the fetus) get is from internal scatter for which there is nothing to do. The controllable scatter mostly comes from the IR and the wall anteriorly.   Some of us have been around long enough to know that we started shielding posteriorly because RT often failed to collimate to the IR back in the bad old days. Posterior shielding was in lieu of good practice, in other words.  When properly collimated, there is very little scatter from the posterior.

LAB DATA:
 


 
Technical/Shielding Factors mR % change in dose
Male gonads 100 kVp, no shielding, collimated 0.8
 
Male gonads 100 kVp, posterior shielding, collimated 0.6 -25%
Male gonads 100 kVp, anterior shielding, collimated 0.5 -37.5%
Male gonads 100 kVp, shielding inside phantom, collimated 0.3 -62.5%
Male gonads 100 kVp, no shielding, open field 10.0 +1250%

 

 

 

 
Female gonads 100 kVp, no shielding, collimated 2.1
 
Female gonads 100 kVp, posterior shielding, collimated 1.7 -19%
Female gonads 100 kVp, anterior shielding, collimated 1.3 -38.1%
Female gonads 100 kVp, shielding inside phantom, collimated 0.8 -62%
Female gonads 100 kVp, no shielding, open field 22.0 +1046%

The data were obtained by use of a 3M chest and abdomen phantom in upright position in front of an chest IR. Dosimeters were placed in the gonad positions. Shielding was 1) not used, 2) used posteriorly, 3) used anteriorly, 4) used internally between chest and abdomen, and 5) no shielding with primary beam overlapping gonads (no collimation).

As you can see, placing a shielding INSIDE the patient does the most good but I would not recommend it. Posterior shielding does some good but not as much as anterior shielding. So, if we can only shield one side, anterior is better but it is always better to over-shield then under-shield.
 

Data kindly supplied by

Dr. Elwin Tilson
Dept. of Radiologic Sciences
Armstrong Atlantic St.Univ
Savannah, GA  31419

www.radsci.armstrong.edu/el/